Operations Manual
BOD
ANALYST
Software That Makes BOD Testing
Easier, More Efficient, and More Accurate
Vue de la page 0
1 2 3 4 5 6 ... 69 70

Résumé du contenu

Page 1 - Operations Manual

Operations ManualBODANALYSTSoftware That Makes BOD TestingEasier, More Efficient, and More Accurate

Page 2

SECTION 2 SAMPLES2.1 WHAT IS A "SAMPLE?"With YSI’s BOD Analyst software you can organize your work by sample. A sample is anythingfor which

Page 3 - CONTENTS

Samples Section 2YSI Incorporated BOD Analyst for Windows Software 7In this example, there are two dilutions: 150ml and 275ml. This Sample is seeded

Page 4

Samples Section 2YSI Incorporated BOD Analyst for Windows Software 8Dilution Setup: You can have up to 5 dilutions for any Sample. On the General Setu

Page 5 - SECTION 1 INTRODUCTION

Samples Section 2YSI Incorporated BOD Analyst for Windows Software 9Quality Control: Used in the trending graph function to display Quality Control Li

Page 6 - STARTING THE PROGRAM

Samples Section 2YSI Incorporated BOD Analyst for Windows Software 10You can find a detailed description of the abbreviation algorithm in the next sec

Page 7 - 1.3 QUICK START GUIDE

SECTION 3 SEEDING SAMPLES3.1 WHY SEED?Seeding is necessary when there are not enough bacteria present in the sample. A small amountof seed material t

Page 8

Seeding Samples Section 3YSI Incorporated BOD Analyst for Windows Software 12If you seed directly to the bottle, the amount of seed material in each b

Page 9

SECTION 4 SAMPLE GROUPS4.1 WHAT IS A "SAMPLE GROUP"?A Sample Group is a set of samples. For brevity, we often refer to Sample Groups as sim

Page 10 - SECTION 2 SAMPLES

Sample groups Section 4YSI Incorporated BOD Analyst for Windows Software 14NOTE: The dilution blank sample is automatically added when you build a bat

Page 11 - 2.3 ADDING A NEW SAMPLE

SECTION 5 BATCHES5.1 WHAT IS A "BATCH?"A Batch is all initial DOs that are done on any one day, along with their corresponding final DOs.Bec

Page 13 -

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 16The Open Batch Form can also be used to delete a batch. Select the batch to be d

Page 14 - 2.6 DELETING A SAMPLE

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 175.4 WORKING WITH BATCHES5.4.1 STEPS TO BUILDING A BATCHAdd the samples for the ba

Page 15 - SECTION 3 SEEDING SAMPLES

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 185.4.5 IMPORT BATCHThis feature allows you to setup a batch by importing a text fi

Page 16

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 19dilution oxygen depletion, seed correction factor and BOD are calculated automati

Page 17 - SECTION 4 SAMPLE GROUPS

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 20Select the sample to add from the list and the number of dilutions and the diluti

Page 18 - 4.5 EDITING A GROUP

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 21Notes: Used for any miscellaneous notes for the entire batch.Bottle Notes: This f

Page 19 - SECTION 5 BATCHES

Batches Section 5YSI Incorporated BOD Analyst for Windows Software 225.14 DELETING A BOTTLETo remove a bottle from a batch, click on therow header and

Page 20 - 5.3 THE BUILD BATCH FORM

SECTION 6 TAKING INITIAL DO READINGS6.1 ENTERING BOTTLE NUMBERS6.1.1 USING THE YSI 5000/5100 DO INSTRUMENTJust after selecting Read / Initials from th

Page 21 - WORKING WITH BATCHES

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 24then press CONFIRM.To enter "4" for the last digit,

Page 22 - 5.5 THE BENCH SHEET

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 25You can also move between dilutions (bottles) without going th

Page 23 - 5.8 ADD A SAMPLE TO THE BATCH

iCONTENTSSECTION 1 INTRODUCTION...11.

Page 24 - 5.12 FIELD DESCRIPTIONS

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 26You do not have to do the final DOs in the same order as you d

Page 25 - 5.13 COLUMN DESCRIPTIONS

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 27The fastest and easiest way is to set Technique: to auto. Aft

Page 26 - 5.14 DELETING A BOTTLE

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 28Now, we will want to measure only one of the dilutions of the

Page 27 - 6.1 ENTERING BOTTLE NUMBERS

Taking Initial DO Readings Section 6YSI Incorporated BOD Analyst for Windows Software 29You don't have to derive initial DOs for every Sample in

Page 28

SECTION 7 TAKING FINAL DO READINGS7.1 SELECTING A BATCHFrom the main bench sheet window, simply enter the sample date for the batch you would like tot

Page 29 - 6.1.2 USING THE BARCODE WAND

Taking Final Dos Readings Section 7YSI Incorporated BOD Analyst for Windows Software 31Your Technique for Auto Stabilize setting affects how the BOD A

Page 30 - 6.2 AUTO STABILIZATION

Taking Final Dos Readings Section 7YSI Incorporated BOD Analyst for Windows Software 327.3 CORRECTING MISTAKESIf you have finished reading the final D

Page 31 - 6.4 DERIVED INITIAL DOS

SECTION 8 REPORTSUnder the File menu, you will find the Print Daily Bench Sheet and Print Everything Reportoptions. The Print Daily Bench Sheet will p

Page 32

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 348.1 REPORT DESIGNThe report designer is used to setup spreadsheet type reports.

Page 33 - 6.5 CORRECTING MISTAKES

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 35Key ActionUp Arrow Move up one rowDown Arrow Move down one rowRight Arrow Move ri

Page 34 - 7.1 SELECTING A BATCH

ii7.3 CORRECTING MISTAKES...32SECTION

Page 35

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 36Note:To resize multiple columns at one time, select all the columns by clicking a

Page 36 - 7.3 CORRECTING MISTAKES

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 373. Release the left mouse button when the pointer is over the new location for th

Page 37 - SECTION 8 REPORTS

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 38ABS(A1)The absolute value of the cell is returned.ISEMPTY(A1)If A1 is empty then

Page 38 - 8.1 REPORT DESIGN

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 39Offset The offset from the start date of the report. For example, if the start

Page 39

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 408.1.9 MONTHLY FUNCTIONSMonthly functions return summary values for a sample for a

Page 40 - 8.1.5 MOVING DATA

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 41The ‘<<< Blank Report >>> selection loads a completely blank

Page 41 - 8.1.7 FUNCTIONS IN FORMULAS

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 428.2 EDIT MENU FUNCTIONSTo cut numbers, data values or text within the currently s

Page 42 - 8.1.7 IF STATEMENTS

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 43The Locate Sample Formula dialog can also be opened by clicking on the Locate Sam

Page 43

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 44Sample Name: Check the sample name box to have the sample name appear as a column

Page 44 - 8.1.9 MONTHLY FUNCTIONS

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 45Display [X] Decimal Places: Enter the number of digits to be displayed after the

Page 45

SECTION 1 INTRODUCTIONBOD Analyst software ™ for Windows™ (BOD Analyst software) is a Windows-based PCsoftware accessory to the YSI 5000 and 5100 Diss

Page 46 - LOCATING DATA ON THE REPORT

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 46Statistic: Sets the summary statistic to use.Decimal Places: The number of digits

Page 47

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 478.4 FORMATTING CELLSThe BOD Analyst software Report Designer offers great flexibi

Page 48

Reports Section 8YSI Incorporated BOD Analyst for Windows Software 488.5 OBTAINING HELP WHILE DESIGNING REPORTSThe help menu provides access to contex

Page 49

SECTION 9 KEEPING YOUR DATA SAFE9.1 DATEThe BOD Analyst software depends on the correct date to do its calculations and manage thedata. The current d

Page 50

Keeping Your Data Safe Section 9YSI Incorporated BOD Analyst for Windows Software 50will be created and opened. You must setup your samples and group

Page 51 - 8.4 FORMATTING CELLS

SECTION 10 GRAPHING10.1 INTRODUCTIONThe graphing function provides the ability to produce time series type (Data vs. Time) graphs.Up to two samples ma

Page 52 - 8.6 PRINTING REPORTS

Graphing Section 10YSI Incorporated BOD Analyst for Windows Software 52Sample 1 and Sample 2 are the fields in which the sample names are selected. If

Page 53 - 9.3 UNDERSTANDING DATA FILES

Graphing Section 10YSI Incorporated BOD Analyst for Windows Software 53differences between printing in color or black and white.10.4 GETTING STATISTIC

Page 54 - 9.5 REPAIR

SECTION 11 CUSTOMIZING YOUR PROGRAMThe BOD Analyst software is flexible enough to handle the wide variety of ways that people dotheir BODs. Some thing

Page 55 - SECTION 10 GRAPHING

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 55number.The auto-stable routine starts immediately after the bot

Page 56 - 10.3 PRINTING A GRAPH

Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 2Click OK to proceed.The installation program will then display the following

Page 57

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 56Maximum acceptable value for Initial DO minus Final DO in the d

Page 58 - 11.2 AUTO-STABILIZATION

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 57The maximum acceptable value of BOD for tests with the sample n

Page 59 - 11.3 THE CRITERIA FORM

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 58want the program to derive. You must measure at least one dilu

Page 60

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 59Technique: This field describes exactly how and when a stable r

Page 61

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 60Initial low temp: If the temperature of your initial DO is belo

Page 62 - 11.5 THE YSI PARAMETERS FORM

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 61Comm port: The program can communicate to the YSI 5000/5100 DO

Page 63

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 62Use Serial Number: Sets whether the program will assign Seria

Page 64

Customizing Your Program Section 11YSI Incorporated BOD Analyst for Windows Software 63Serial Number forRAW 10 bottle

Page 65 - 11.7 SERIAL NUMBERS

SECTION 12 CONTACT INFORMATIONFor any questions regarding PC6000 software, or any YSI equipment, please contact YSITechnical Support at (800) 897-4151

Page 66

SECTION 13 ACCESSORIESINSTRUMENTS! 5000-115 Oxygen Instrument, Digital, Lab-115V! 5100-115 Oxygen Instrument, Digital, Lab with SOUR software – 115VPR

Page 67 - RAW 10 bottle

Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 3The Bench sheet is the main form of the BOD Analyst software. From the bench

Page 68

1700/1725 Brannum LaneYellow Springs, Ohio 45387 USA(800) 765-4974 (937) 767-7241FAX: (937) 767-9320Website: http://www.ysi.comE-mail: [email protected]

Page 69 - SECTION 13 ACCESSORIES

Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 4Volume of bottles (ml) 300Dilution Unit mlSeed Added to BottleReport Font Ari

Page 70

Introduction Section 1YSI Incorporated BOD Analyst for Windows Software 5Coefficient of Variation Limit 0.35Apply Criteria after Final Readings OnStep