YSI 5200A
59
YSI 5200A
58
Network configurations
Master
Master
Master
slave
slave
slave
slave
slaveslave slave slave
slave slave
slave
slaveslaveslave
slave
do not wire “star” network
follow master/slave wiring directions figure 3.26
Master slave
slave
slave slave
do not wire “loop” network
b
b
good
good
bad bad
Figure 3.28
Test Network Wiring and Conguration
Apply power to only the master and last slave device on the network. If the master is
not the rst physical device on the network, apply power to the master and the rst
and last physical slave devices on the network. See Network beginning on page 95.
Step 8 Wire Communication Equipment
Communication equipment must be installed in order to communicate, either lo-
cally or remotely, with the 5200A. e communication connection can be made
locally (RS232) or remotely with a serial to ethernet device. Only one connection
method can be used at a time. AquaManager supports both connection methods.
Installation and Wiring
Notes:
- Firmware can only be uploaded to an instrument using a local RS232 con-
nection. See Update 5200A Firmware - page 112.
- On a network, install and congure RS232 or the serial to ethernet device
hardware on the master node only.
Email Alarming
Email alarming can be congured in one of two ways:
- from a PC running the AutoPoll Program that is included with AquaMan-
ager soware (PC must have internet connection). is is referred to as
AquaManager email.
- via a serial to ethernet* device with TCP/IP connection at master 5200A.
is is referred to as 5200A email.
*Serial to ethernet connection requires access to a SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location H - gure 3.29 of I/O
board to PC using RJ45 to DB9 adapter.
Directions to Wire Direct RS232 PC Connection
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT: déconnectez lalimentation externe de l’unité avant deectuer
un câblage quelconque.
1. Perform steps 1 - 4 of 5200A installation and steps 5, 6 and 7 as required
for the application.
2. Open front panel - page 39.
3. Feed direct connect cable through drilled hole of rubber grommet and
to location H.
4. Run high and low voltage cables through separate bulkhead and con-
duit.
5. Connect one end of communication cable to RJ45 at location H. Connect
other end to PC com port using RJ45 to DB9 adapter.