YSI 5200A
69
YSI 5200A
68
Sokey label Function
Exit menu and return to Run Screen saving
any menu conguration changes.
Exit menu (to previous menu) without sav-
ing conguration changes.
Back to previous menu. Changes to congu-
rations are saved at return to Run Screen.
Prompt to take next action, i.e. enter a value
at calibration menus.
Save alpha and numeric congurations and
calibrations.
Disable/Enable menu option, i.e. buzzer.
Scroll le and right in alpha and numeric
keypad screens.
To acknowledge and save sensor calibration.
Prompt to conrm “Are you sure?” message
when performing certain functions, i.e. reset
to factory and clear data log menus.
Digital input control conguration from
Sensor Setup Menu.
Table 4.1
Run Screen
e Run Screen displays measurements and inputs from enabled sensors and condi-
tions - gure 4.2. At power up, the Run Screen is displayed immediately aer the
YSI splash screen (the YSI splash screen appears for approximately 2 seconds). Icons
displayed at the Run Screen represent individual and systemwide conditions. See
Run Screen Icons - page 70.
Run Screen and sensor display formats are user selectable. See Display Settings -
page 111 and Sensor Setup Menus - page 131.
Notes:
- e Run Screen must be displayed during normal operation. All alarm
functions (5200A alarm emails, energizing of enabled alarm relays,
alarm icons, and buzzer) are suspended when in 5200A instrument
menus. Alarm systems are reset when menus are exited to Run Screen.
Conguring the 5200A
- Always manually poll the 5200A aer uploading conguration changes
and verify the uploaded information is retained. When conguration
changes are uploaded using AquaManager at the same time conguration
changes are being made at the instrument, the AquaManager uploads
may not be saved depending on the timing of keystrokes at the 5200A.
- e Run Screen must be displayed during normal operation. If the
5200A is not at the Run Screen during AquaManager uploading, press-
ing
at the 5200A will only save certain conguration changes.
- At power up, 5200A systems have a 30 second hold o allowing sensor
values to stabilize before enabling any control and/or alarm devices.
Figure 4.2
Conguring the 5200A