5200A slave alarms will be sent. erefore, It is critical that all alarm
emails are addressed locally by the user.
Below is an example of a 5200A DO Alarm Low email sent from AquaManager. All
of the information is in the body of the email and not the subject line.
YSI5200A Alarm
slave 2 (node name in Aquamanager tree)
53-2345567
DO Low Alarm: 68
Notes on AquaManager Alarm Emails:
- AquaManager emails are retransmitted based on the Email Retry Delay
Conguration see page 99.
- Multiple system alarms are included in one AquaManager email.
- AquaManager DO alarm emails provide the %Sat value and do not pro-
vide the mg/L value.
- For the AquaManager alarm email to function properly, it may be neces-
sary to send a test email to each email address congured in the Email
Alarm List in the Node Properties window. Test emails can be sent in the
AquaManager Properties window under the Autopoller tab.
Ethernet
Menus → System → Communications → Ethernet
e serial to ethernet device speed and format are self congured at power up (TCP/
IP connections). If the serial to ethernet device is not installed or is not installed
properly, the ethernet menu will display (Not Available).
Notes:
- 5200A email alarming from a serial to ethernet device requires access to
an SMTP server accessible by the local network.
- Do not recongure internal serial to ethernet device settings.
Conguring the 5200A
1. Install serial to ethernet device according to instruction provided by YSI
with the module.
2. e Ethernet icon
displays at the Run Screen when the serial to ethernet
device is properly installed.
3. Enter Ethernet menu:
Menus → System → Communications → Ethernet
4.
Use
and
to scroll and highlight submenus.
5.
Press
to select submenu.
6. Congure submenus as needed.
7. Press
to save conguration and return to Run Screen.
Notes:
- e CC/Email address must be congured when using 5200A email alarm-
ing. Specic alarm system information will not be sent if the CC/Email
address menu is not congured with a valid email address.
- If Autopoller is being used for AquaManager email alarming on a TCP/
IP connection, it is strongly recommended that the 5200A email alarming
using the serial to ethernet device be disabled. When the ethernet device
sends alarm emails, all Autopoller SMS communication is suspended. is
could potentially delay any AquaManager alarm emails from being sent.
- To disable the serial to ethernet device from sending alarms, clear all four
email addresses from the Communication  Ethernet menus. Clear the
email addresses by using the “Spc” (space) key on the numeric keypad.
Save change at numeric keypad and press
to save conguration
and return to Run Screen.
DHCP
Enable DHCP if you are using the serial to ethernet device only for 5200A email
alarming and not for AquaManager remote TCP/IP access.