OPERATIONS MANUALba76157e01 03/2014TruLab pH/ISE 1320(P)PH/ISE METERpH/ISE 1320PpH/ISE 1320PrintermodulpH/ISE 1320PpH/ISE 1320
Safety TruLab pH/ISE 1320 10 ba76157e01 03/20142.2 Safe operation2.2.1 Authorized useThis meter is authorized exclusively for pH, ORP and ISE concentr
Glossary TruLab pH/ISE 1320 100 ba76157e01 03/2014Measured value The measured value is the special value of a measured parameter to be determined. It
TruLab pH/ISE 1320 Indexba76157e01 03/2014 10117 IndexAAutomatic switch-off . . . . . . . . . . . . . . . . . . 20AutoReadORP . . . . . . . . . . . .
Index TruLab pH/ISE 1320 102 ba76157e01 03/2014TTemperature measurementISE . . . . . . . . . . . . . . . . . . . . . . . . . . . .47pH . . . . . . .
TruLab pH/ISE 1320 Contact Information103ba76157e01 03/201418 Contact Information18.1 Ordering & Technical SupportWhen placing an order please hav
Contact Information TruLab pH/ISE 1320104ba76157e01 03/2014
1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com
TruLab pH/ISE 1320 Commissioningba76157e01 03/2014 113 Commissioning3.1 Scope of delivery pH meter TruLab pH/ISE 1320 4 batteries 1.5 V Mignon type
Commissioning TruLab pH/ISE 1320 12 ba76157e01 03/20143.3.1 Inserting the batteriesYou can operate the meter either with normal batteries or with rech
TruLab pH/ISE 1320 Commissioningba76157e01 03/2014 133.3.2 Connecting the power pack3.3.3 Mounting the standThe stand base can be mounted at the right
Printer (only TruLab pH/ISE 1320 P) TruLab pH/ISE 1320 14 ba76157e01 03/20144 Printer (only TruLab pH/ISE 1320 P)4.1 Commissioning / switching the pri
TruLab pH/ISE 1320 Printer (only TruLab pH/ISE 1320 P)ba76157e01 03/2014 154.2 Operation / printingData are only output to the printer if the followin
Printer (only TruLab pH/ISE 1320 P) TruLab pH/ISE 1320 16 ba76157e01 03/20144.5 What to do if... / printer Integrated printer does not print Printer o
TruLab pH/ISE 1320 Operationba76157e01 03/2014 175Operation5.1 General operating principles5.1.1 KeypadIn this operating manual, keys are indicated by
Operation TruLab pH/ISE 1320 18 ba76157e01 03/20145.1.2 Display5.1.3 Status information (meter)1 Status information (sensor)2 Measured value (with uni
TruLab pH/ISE 1320 Operationba76157e01 03/2014 195.1.4 Socket field5.2 Switching on the meter1 pH electrode or ion selective electrode 12 Reference el
2 ba76157e01 03/2014TruLab pH/ISE 1320For the most recent version of the manual, please visit www.ysi.com.Contact YSI 1725 Brannum Lane Yellow Springs
Operation TruLab pH/ISE 1320 20 ba76157e01 03/20145.3 Switching offAutomatic switch-off The instrument has an automatic switch-off function in order t
TruLab pH/ISE 1320 Operationba76157e01 03/2014 21Only those displays and functions are available in the active operating mode that are currently being
Operation TruLab pH/ISE 1320 22 ba76157e01 03/2014 Settings Settings are indicated by a colon. The current setting is displayed on the right-hand sid
TruLab pH/ISE 1320 Operationba76157e01 03/2014 235.4.5 Example 1 on navigation: Setting the language1. Press the <On/Off> key. The measured valu
Operation TruLab pH/ISE 1320 24 ba76157e01 03/20145. Select the General submenu with <><>. The current selection is displayed with a fr
TruLab pH/ISE 1320 Operationba76157e01 03/2014 255.4.6 Example 2 on navigation: Setting the date and timeThe measuring instrument has a clock with a d
Operation TruLab pH/ISE 1320 26 ba76157e01 03/20147. If necessary, set the Date and Date format. The setting is made similarly to that of the time.8.
TruLab pH/ISE 1320 Operationba76157e01 03/2014 275.5 Channel displayThe TruLab pH/ISE 1320 manages the connected sensors and displays which sensor is
pH value TruLab pH/ISE 1320 28 ba76157e01 03/20146 pH value6.1 Measuring6.1.1 Measuring the pH valueNOTE When connecting an earthed PC/printer, measur
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 29Criteria for a stable measured valueThe Stability control function checks whether the measured values
TruLab pH/ISE 1320 - ContentsTruLab pH/ISE 1320 Contentsba76157e01 03/2014 31 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH value TruLab pH/ISE 1320 30 ba76157e01 03/20146.1.2 Measuring the temperatureFor reproducible pH measurements, it is essential to measure the tempe
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 316.2 Calibration6.2.1 Why calibrate?pH electrodes age. This changes the zero point (asymmetry) and slop
pH value TruLab pH/ISE 1320 32 ba76157e01 03/2014Continuing with two-point calibration5. Thoroughly rinse the electrode with deionized water.6. Immers
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 33Continuing with three- to five-point calibration14. Start the measurement with <ENTER>.The measu
pH value TruLab pH/ISE 1320 34 ba76157e01 03/20146.2.4 Manual calibration (ConCal)Make sure that in the sensor menu, Buffer menu, the ConCal buffer se
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 35Continuing with two-point calibration7. When measuring without temperature sensor:Enter the temperatur
pH value TruLab pH/ISE 1320 36 ba76157e01 03/2014Continuing with three- to five-point calibration14. Immerse the electrode in the second buffer soluti
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 376.2.5 Calibration pointsCalibration can be performed using one to five buffer solutions in any order (
pH value TruLab pH/ISE 1320 38 ba76157e01 03/20146.2.6 Calibration dataDisplaying the calibration data The calibration record of the last calibration
TruLab pH/ISE 1320 pH valueba76157e01 03/2014 39Calibration record (example)Display Calibration recordZero point [mV]Slope [mV/pH]+++- 15 ... + 15 -6
Contents TruLab pH/ISE 1320 4 ba76157e01 03/20145.4.4 Elements in menus and dialogs . . . . . . . . . . . . . . . . . . . 215.4.5 Example 1 on navig
pH value TruLab pH/ISE 1320 40 ba76157e01 03/20146.2.7 Continuous measurement control (CMC function) The Continuous Measurement Control (CMC funct
TruLab pH/ISE 1320 ORP voltageba76157e01 03/2014 417 ORP voltage7.1 Measuring7.1.1 Measuring the ORPNOTE When connecting an earthed PC/printer, measur
ORP voltage TruLab pH/ISE 1320 42 ba76157e01 03/2014The minimum duration until a measured value is assessed as stable is the monitored time interval.
TruLab pH/ISE 1320 ORP voltageba76157e01 03/2014 437.1.2 Measuring the relative ORPTo measure the difference of the ORPs of two solutions, you have to
ORP voltage TruLab pH/ISE 1320 44 ba76157e01 03/2014The minimum duration until a measured value is assessed as stable is the monitored time interval.
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 458 Ion concentration8.1 Measuring8.1.1 Measuring the ion concentration NOTE When connecting an
Ion concentration TruLab pH/ISE 1320 46 ba76157e01 03/2014 as soon as the measured value is outside the stability range when the automatic Stabilit
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 478.1.2 Measuring the temperatureFor reproducible ion-selective measurements, it is essential t
Ion concentration TruLab pH/ISE 1320 48 ba76157e01 03/20148.2 Calibration8.2.1 Why calibrate?Ion-selective electrodes age and are temperature-dependen
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 49Continuing with two-point calibration7. Thoroughly rinse the electrode with distilled water.8
TruLab pH/ISE 1320 Contentsba76157e01 03/2014 59.3 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . . . . . . 759.3.1 System .
Ion concentration TruLab pH/ISE 1320 50 ba76157e01 03/2014 Continuing with three- to seven-point calibrationRepeat the steps 8 to 13 in the same way w
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 518.2.4 Calibration standardsUse two to seven different standard solutions. The standard soluti
Ion concentration TruLab pH/ISE 1320 52 ba76157e01 03/2014Calibration evaluationAfter calibrating, the meter automatically evaluates the calibration.C
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 538.3 Selecting the measuring methodThe following methods are supported: Standard addition St
Ion concentration TruLab pH/ISE 1320 54 ba76157e01 03/20148.3.1 Standard additionIn the "Standard addition" procedure, a known amount of sta
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 5511. Select Continue with <> <> and start the measurement with <MENU/ENTER>
Ion concentration TruLab pH/ISE 1320 56 ba76157e01 03/20148.3.2 Standard subtractionIn the "Standard Subtraction" procedure, a known amount
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 5710. Using <> <> and <MENU/ENTER>, select the values for the vol-ume of th
Ion concentration TruLab pH/ISE 1320 58 ba76157e01 03/201419. Confirm Yes with <MENU/ENTER>. Measurement with the selected method is completed.
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 598.3.3 Sample additionIn the "Sample addition" procedure, a known amount of sample i
Contents TruLab pH/ISE 1320 6 ba76157e01 03/2014
Ion concentration TruLab pH/ISE 1320 60 ba76157e01 03/201411. Select Continue with <> <> and start the measurement with <MENU/ENTER&g
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 618.3.4 Sample subtractionWith the "Sample subtraction" procedure, a known amount of
Ion concentration TruLab pH/ISE 1320 62 ba76157e01 03/2014Measuring7. Start the measurement with <F1> Continue. An entry window appears.8. Prepa
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 638.3.5 Standard addition with blank value correction (Blank value addition) In the "S
Ion concentration TruLab pH/ISE 1320 64 ba76157e01 03/2014MeasuringCALIBRATION12.12.2012 12:12:12#1 0.010 mg/l 20.2 mV 25.0 °C#2 0.100 mg/l 79.2 mV
TruLab pH/ISE 1320 Ion concentrationba76157e01 03/2014 6512. Add the standard solution to the sample.13. Using <> <> and <MENU/ENTER&
Settings TruLab pH/ISE 1320 66 ba76157e01 03/20149 Settings9.1 Measurement settings9.1.1 Settings for pH measurementsThe settings for pH measurements
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 67Calibration/ Serial number (sen-sor)- Entry of the series number of the connected sensor. The series n
Settings TruLab pH/ISE 1320 68 ba76157e01 03/20149.1.2 Buffer sets for calibrationYou can use the buffer sets quoted in the table for an automatic cal
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 69No. Buffer set * pH values at1 YSI * 4.0007.00010.00025 °C2 ConCal Any Any3 NIST/DINDIN buffers accord
TruLab pH/ISE 1320 Overviewba76157e01 03/2014 71 Overview1.1 TruLab pH/ISE 1320 meterThe compact precision pH meter TruLab pH/ISE 1320 enables you to
Settings TruLab pH/ISE 1320 70 ba76157e01 03/201413 Fisher * 2.0074.0027.00410.00225 °C14 Fluka BS * 4.0066.9848.95725 °C15 Radiometer * 1.6784.0057
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 71* Brand names or trade names are trademarks of their respective owners protected by law.9.1.3 Calibrat
Settings TruLab pH/ISE 1320 72 ba76157e01 03/20149.1.4 Settings for ORP measurementsThe settings for ORP measurements are made in the menu for calibra
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 739.2 Settings for ISE measurementsThe settings are made in the measuring menu of the ISE measurement. T
Settings TruLab pH/ISE 1320 74 ba76157e01 03/2014Man. temperature -25 ... +25 ... +130 °C Entry of the manually deter-mined temperature. For measureme
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 759.3 Sensor-independent settings9.3.1 SystemTo open the Storage & config menu, press the <MENU/E
Settings TruLab pH/ISE 1320 76 ba76157e01 03/20149.3.2 Data storageThis menu contains all functions to display, edit and erase stored measured values
TruLab pH/ISE 1320 Settingsba76157e01 03/2014 77The sensor settings are reset under the Reset menu item in the menu for cali-bration and measurement s
Data storage TruLab pH/ISE 1320 78 ba76157e01 03/201410 Data storageYou can transmit measured values (datasets) to the data storage: Manual data stor
TruLab pH/ISE 1320 Data storageba76157e01 03/2014 79You have the following options: To erase the entire storage, confirm Yes. To cancel the storage
Overview TruLab pH/ISE 1320 8 ba76157e01 03/20141.2 TruLab pH/ISE 1320P meterThe integrated printer of the TruLab pH/ISE 1320 P enables to document me
Data storage TruLab pH/ISE 1320 80 ba76157e01 03/2014Settings You can configure the automatic data storage function with the following settings:Starti
TruLab pH/ISE 1320 Data storageba76157e01 03/2014 81Energy saving mode ([Eco mode])If the automatic storing function is active, the meter provides an
Data storage TruLab pH/ISE 1320 82 ba76157e01 03/2014SettingsDisplay presentation of a datasetRepresentation of a dataset (USB output)Menu item Settin
TruLab pH/ISE 1320 Data storageba76157e01 03/2014 83Quitting the display To quit the display of stored measurement datasets, you have the following op
TruLab pH/ISE 1320 Transmitting data (USB interface)ba76157e01 03/2014 8411 Transmitting data (USB interface) 11.1 Options for data transmissionVia th
TruLab pH/ISE 1320 Transmitting data (USB interface)ba76157e01 03/2014 8511.2 Connecting a PC Connect the TruLab pH/ISE 1320 to the PC via the USB in
Transmitting data (USB interface) TruLab pH/ISE 1320 86 ba76157e01 03/2014
TruLab pH/ISE 1320 Maintenance, cleaning, disposalba76157e01 03/2014 8712 Maintenance, cleaning, disposal12.1 Maintenance12.1.1 General maintenance ac
Maintenance, cleaning, disposal TruLab pH/ISE 1320 88 ba76157e01 03/201412.2 CleaningOccasionally wipe the outside of the measuring instrument with a
TruLab pH/ISE 1320 What to do if...ba76157e01 03/2014 8913 What to do if...13.1 pH/ORPError message OFL, UFLThe measured value is outside the measurin
TruLab pH/ISE 1320 Safetyba76157e01 03/2014 92 Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual provi
What to do if... TruLab pH/ISE 1320 90 ba76157e01 03/2014No stable measured value Obviously incorrect measured valuesCause RemedypH electrode:– Juncti
TruLab pH/ISE 1320 What to do if...ba76157e01 03/2014 9113.2 ISEError message OFL Obviously incorrect measured valuesError message, Error (invalid cal
What to do if... TruLab pH/ISE 1320 92 ba76157e01 03/2014
TruLab pH/ISE 1320 What to do if...ba76157e01 03/2014 9313.3 General informationSensor symbol flashes displayMeter does not react to keystrokeYou want
Technical data TruLab pH/ISE 1320 94 ba76157e01 03/201414 Technical data 14.1 Measuring ranges, resolution, accuracy14.1.1 pH/ORPMeasuring ranges, re
TruLab pH/ISE 1320 Technical databa76157e01 03/2014 9514.1.2 ISEMeasuring ranges, resolutionParameter Measuring range ResolutionISE [mg/l] 0.000 ...
Technical data TruLab pH/ISE 1320 96 ba76157e01 03/201414.2 General dataDimensions TruLab pH/ISE 1320: Approx. 230 x 190 x 80 mmTruLab pH/ISE 1320 P:A
TruLab pH/ISE 1320 Technical databa76157e01 03/2014 97Applicable directives and standardsEMC EC directive 2004/108/ECEN 61326-1EN 61000-3-2EN 61000-3-
Firmware update TruLab pH/ISE 1320 98 ba76157e01 03/201415 Firmware updateAvailable firmware updates are provided on the Internet. With the firmware u
TruLab pH/ISE 1320 Glossaryba76157e01 03/2014 9916 GlossarypH/ORP/ISEGeneral informationAsymmetry see zero pointElectromotive force of a electrodeThe