Ysi Pro1030 Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 21
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
26
27
1. Soak the sensor for 10-15 minutes in clean water containing a
few drops of commercial dish washing liquid.
2. GENTLY clean the glass bulb and platinum button by rubbing
with a cotton swab soaked in the cleaning solution.
3. Rinse the sensor in clean water, wipe with a cotton swab
saturated with clean water, and then rerinse with clean water.
If good pH and/or ORP response is still not restored, perform the
following additional procedure:
1. Soak the sensor for 30-60 minutes in one molar (1 M) hydrochloric
acid (HCl). This reagent can be purchased from most lab supply
distributors. Be sure to follow the safety instructions included
with the acid.
2. Rinse the sensor in clean water, wipe with a cotton swab
saturated with clean water (not DI water), and then rerinse with
clean water. To be certain that all traces of the acid are removed
from the sensor crevices, soak the sensor in clean water for
about an hour with occasional stirring.
If biological contamination of the reference junction is suspected or if
good response is not restored by the above procedures, perform the
following additional cleaning step:
1. Soak the sensor for approximately 1 hour in a 1:1 dilution of
commercially-available chlorine bleach.
2. Rinse the sensor with clean water and then soak for at least 1
hour in clean water with occasional stirring to remove residual
bleach from the junction. (If possible, soak the sensor for a
period of time longer than 1 hour in order to be certain that all
traces of chlorine bleach are removed.) Then rerinse the sensor
with clean water and retest.
CAUTION: When using a cotton swab, be careful NOT to
wedge the swab between the guard and the glass sensor. If
necessary, remove cotton from the swab tip, so that the cotton
can reach all parts of the sensor tip without stress. You can
also use a pipe cleaner for this operation if more convenient.
i
Dry the port and sensor connector with compressed air and
apply a very thin coat of o-ring lubricant to all o-rings before
reinstallation.
i
If this procedure is unsuccessful, as indicated by improper sensor
performance, contact YSI Technical Support or the YSI authorized
dealer where you purchased the instrument.
SENSOR STORAGE
SHORT TERM STORAGE
The instrument is supplied with a grey storage sleeve that slides over
the probe guard. The sleeve is used for short-term storage (less than 2
weeks). Be sure to keep a small amount of moisture (clean tap water)
on the sponge in the sleeve during storage. The moistened sponge in
the sleeve provides a 100% water saturated air environment which is
ideal for short-term sensor storage.
LONG TERM STORAGE
The conductivity sensor should be stored long term in a dry state while
the ISE sensor should be stored in solution. When storing for more then
30 days, place the ISE sensor in the storage bottle that was originally
included with the sensor. This can be filled with buffer 4 solution. If you
no longer have the storage bottle, simply place the sensor in a buffer 4
solution. Ensure the conductivity sensor is clean and dry.
Long Term Storage Temperature: -5 to 70°C (23 to 158°F) without pH
0 to 30°C (32 to 86°F) with pH*
*Operating temperature range for pH sensor is -5 to 60°C (23 to 140°C).
TROUBLESHOOTING
Symptom Possible Solution
Instrument will not turn
on, a battery symbol
appears, or “Critical
Shutdown” displays
on the screen.
1. Low battery voltage, replace batteries.
2. Batteries installed incorrectly,
check battery polarity.
3. Return system for service.
Temperature values
display Over or Undr
on Run screen.
1. Sample temperature is less than -5° C or
more than +55°C. Increase or decrease
the sample temperature to bring within
the allowable range.
2. Contact YSI Tech Support.
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21

Commentaires sur ces manuels

Pas de commentaire