Ysi Pro1030 Manuel d'utilisateur Page 16

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 21
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 15
24
25
13, and 15 becomes 14. The instrument will return to Data mode after
erasing one data set.
Select Erase All Data Sets and press enter to clear the Pro1030 memory
and return to Data mode.
EXITING DATA MODE
While in Data mode, press enter to access the bottom functions. Next,
highlight the ESC-Exit box and press enter to return to the Run screen.
CARE, MAINTENANCE AND
STORAGE
This section describes the proper procedures for care, maintenance
and storage of the sensors. The goal is to maximize their lifetime and
minimize down-time associated with improper sensor usage.
GENERAL MAINTENANCE
GENERAL MAINTENANCE - GASKET AND O-RINGS
The instrument utilizes a gasket and o-rings as seals to prevent
water from entering the battery compartment and the sensor port.
Following the recommended procedures will help keep the instrument
functioning properly.
If the gasket, o-rings and sealing surfaces are not maintained properly,
it is possible that water can enter the battery compartment and/or
sensor port of the instrument. If water enters these areas, it can damage
the battery terminals or sensor port causing loss of battery power, false
readings and corrosion to the sensors or battery terminals. Therefore,
when the battery compartment lid is removed, the gasket that provides
the seal should be carefully inspected for contamination (i.e. debris,
grit, etc.) and cleaned with water and mild detergent if necessary.
The same inspection should be made of the o-rings associated with
the ISE sensor connector when replacing the ISE sensor. The o-rings
should be free of dirt or debris before installing the sensor onto the
cable.
GENERAL MAINTENANCE - ISE SENSOR PORT
It is important that the entire sensor connector end be dry when
installing, removing or replacing the sensor. This will prevent water
from entering the port. Once the ISE sensor is removed, examine the
connector inside the port. If any moisture is present, use compressed
air to completely dry the connector or let it air dry. If the connector is
corroded, contact YSI Technical Support or the YSI authorized dealer
where you purchased the instrument.
SENSOR MAINTENANCE
SENSOR MAINTENANCE - TEMPERATURE
You must keep the temperature sensor free of build up. No additional
maintenance is required. A toothbrush can be used to scrub the
temperature sensor if needed.
SENSOR MAINTENANCE - CONDUCTIVITY
The openings that allow sample access to the conductivity electrodes
should be cleaned regularly. The small cleaning brush included in
the Maintenance Kit is intended for this purpose. Dip the brush in
clean water and insert it into each hole 10 to 12 times. In the event
that deposits have formed on the electrodes, it may be necessary to
use a mild detergent (laboratory grade soap or bathroom foaming tile
cleaner) with the brush. Rinse thoroughly with clean water, then check
the response and accuracy of the conductivity cell with a calibration
solution.
SENSOR MAINTENANCE - pH AND ORP
Cleaning is required whenever deposits or contaminants appear on the
glass and/or platinum sensor surfaces or when the sensor’s response
slows. The cleaning can be chemical and/or mechanical.
Removing the sensor from the cable may make cleaning easier.
Initially, use clean water and a soft clean cloth, lens cleaning tissue, or
cotton swab to remove all foreign material from the glass bulb and/
or platinum button. Then use a moistened cotton swab to carefully
remove any material that may be blocking the reference electrode
junction of the sensor.
If good pH and/or ORP response is not restored, perform the following
additional procedure:
Typical working life for pH and ORP sensors is approximately
12-24 months depending on usage, storage and
maintenance. Proper storage and maintenance generally
extends the sensor’s working life.
i
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21

Commentaires sur ces manuels

Pas de commentaire